    • 商品名称: F97XP domestic fluorescence spectrometer

     

    technical parameter

    Technical indicators

    Instrument model

    F97XP

    bandwidth

    excitation

    1nmoptional*/2nm/5nm/10nm/20nm

    emission

    light source

    High-strength and long-life Japan Hamasong imported 150W automatic deodorant xenon lamp

    optical system

    Double Monochromator

    Excitation wavelength range

    200nm~900nm

    Emission wavelength range

    200nm~900nm

    data transmission method

    USB2.0interface

    power supply

    220V/50Hz110V/60Hz

    Standard energy consumption

    200W

    Dimensions (mm)

    380×445×310

    weight

    Gross weight 14kg, net weight 12kg

  •  

    Standard configuration items

    quantity

    Fluorescence spectrophotometer host

    one  tower

    Instruction Manual

    one  book

    Product Qualification Certificate

    one serving

    Application References

    one serving

    Software U Disk

    one

    10mm Stone cuvettes

    2     only

     power cable

    one  root

    USB cable data line

    one  root

    fuse(2A 5A

    2 each

    Packing List

    one serving

    dust cover

    one piece

  •  

    Measurement accessories

    Function

    Single-hole sample cell holder (already standard)

    Routine Liquid Fluorescence Measurements

    Multipurpose fluorescence sample holder

    Various measurement accessory bases

     Frequency doubling filter

    Eliminate frequency doubling interference peaks in fluorescence measurements

    200μLMicrocentrifuge Tube Measurement Accessories

    Centrifuge tube fluorescence measurement function

    Capillary Micro Sample Measurement Accessories

    Provides capillary micro sample measurement function

    Single Well Cell Adapter

    Combined with the multi-purpose fluorescence sample holder to form the functions of conventional liquid fluorescence measurement and fluorescence reflection measurement

    Membrane fluorescence sample measuring rack

    Combined with the multi-purpose fluorescent sample holder to form conventional film sample fluorescence measurement and fluorescence reflection and transmission measurement functions

    Powder Fluorescence Sample Measuring Rack

    Combined with the multi-purpose fluorescence sample holder to form a conventional powder sample fluorescence measurement function

    Sheathed cell holder

    Combined with the multi-purpose fluorescence sample holder to form a sheathed fluorescence sample cell with fluorescence measurement function

    Micro Quartz Four-way Cuvette

    Provides fluorescence measurement function for trace samples

  • • Double monochromator optical design, key components such as long-life ozone-removing xenon lamp light source and photomultiplier tube detector used in the instrument are all original components from Hamamatsu, Japan, which ensures that the instrument can obtain sufficient energy in the spectral range of 200nm-900nm. Light intensity signal and detection sensitivity.

    • The wavelength scanning speed can reach up to 48000nm/min, which can easily realize the three-dimensional fluorescence spectrum scanning of the sample within 1 minute.

    • With multi-level bandwidth adjustable function (F97 is a fixed 10nm bandwidth).

    • It can realize fluorescence three-dimensional scanning function, synchronous scanning function of equal wavelength difference, synchronous scanning function of equal wave number difference (constant energy difference), etc. It is an ideal instrument for researchers to conduct three-dimensional fluorescence analysis and synchronous fluorescence analysis. (*Synchronous scan function is only available on F97XP)

    • A variety of measurement accessories: including film sample and powder sample measurement accessories, up-conversion fluorescence measurement accessories, fluorescent sample semi-automatic sampling accessories, semiconductor constant temperature sample cell accessories, water bath temperature control cuvette rack accessories, etc. Rich accessories greatly expand the application range of the instrument.

    • The built-in programmable optical gate device is specially designed for the measurement of samples that are prone to light reaction or the fluorescence signal is unstable, and improves the accuracy of fluorescence analysis by unifying the sample excitation and detection rhythm.

    • Spectral correction function: The spectral correction function can correct the spectral response function of the instrument itself and obtain the real fluorescence spectrum data of the sample.

    • Optional audit trail function (with multi-level management authority), 3Q certification service.

  • Instrument introduction:
      F97 series fluorescence spectrophotometers (F97, F97Pro, F97XP) are high-end fluorescence spectrophotometer products carefully developed by Shanghai Prism Technology Co., Ltd. The product adopts dual monochromator, proportional dual optical path design with excitation light monitoring system, 150W Hamamatsu high-quality xenon lamp, adopts 1200 lines/mm concave grating and large aperture aspheric mirror spectroscopic system, small size, compact structure, and has the ability to detect It has the characteristics of high sensitivity, fast scanning speed, wide spectral measurement range, large detection dynamic range and fast three-dimensional scanning. Excellent mechanical structure and digital signal processing ensure that the instrument can complete 3D full-spectrum scanning within 1 minute during high-speed 3D scanning.
    F97 series fluorescence spectrophotometers (F97, F97Pro, F97XP) easily meet the qualitative and quantitative analysis needs in the fields of material research, drug analysis, biochemical and clinical testing, water quality analysis and control, food safety testing and other fields. It is your best choice for multi-functional, high-precision, high-efficiency fluorescence analysis!